DxI 9000 Access Immunoassay Analyzer

Analytical method

  • Chemiluminescent detector: luminometer

Barcoded reagents

  • Automatic tracking of:
    • Number of tests
    • Available tests
    • Expiration date
    • Lot number
    • Calibration expiration and printouts

Calibration

  • Calibration curve stability up to 64 days (assay dependent)
  • Calibration curves and parameters displayed on screen

Communication modes

  • Unidirectional, bi-directional, bi-directional with true host query
  • Both RS-232 and LAN connections are supported

Compartment temperatures

  • Incubator and wash/read wheel: 98.6°F (37°C)
  • Sample wheel: 40.1°F to 57.2°F (4.5°C to 14°C)
  • Reagent compartment: 39.2°F to 50°F (4°C to 10°C)

Immunoassay menu capacity

  • >50 preprogrammed, barcoded immunoassay methods currently available

Measurement principles

  • Acridan based chemiluminescent Lumi-Phos PRO

Sample container sizes

  • Ability to accommodate any tube between 12–16mm in diameter and 75–100mm in height
  • Sample cups:
    • 0.5 mL
    • 2.0 mL
    • 3.0 mL
  • 2.0 mL insert cups
  • Pediatric insert
  • Auto aliquot tube

Sample management/capacity

  • 140 sample tubes (20 racks of 7 tubes), unique continuous loading feature
  • Primary tube released after aliquot is made (approx. 1 minute) 300 aliquots (≤200 μL) per hour
  • Capable of automation connection
  • Barcode symbology:
    • Ability to read 1-D and 2-D symbologies with camera technology.
  • Sample volume:
    • 2-100 μL (assay dependent)
    • 500 μL in aliquot vessel, maximum (including any potential reflex tests)
  • Sample and reagent pipettor obstruction detection

Sample types (assay dependent, partial list)

  • Serum
  • Plasma
  • Urine
  • Amniotic fluid
  • Whole blood

Reagent storage capacity

  • 50 reagent packs:
    • System menu can be configured to match laboratory testing patterns (ability to load one pack of 50 different assays or 50 packs of one assay)
  • Automatic disposal of empty packs

Throughput

  • Up to 450 tests/hr maximum (one-step assays)

Installation Requirements

Dimensions and weight

  • Analyzer:
    • Weight: 1,785 lbs (810 kg)
    • Length: 79 in (200 cm)
    • Height: 63 in (159 cm)
    • Depth: 41 in (104 cm)
  • Clearance Requirements:
    • Clearance around the analyzer for ventilation
    • Rear: Minimum of 20 in (50 cm)
  • Clearance around the analyzer for service:
    • Rear: Minimum of 20 in (50 cm)
    • Left: 0 in (0 cm)
    • Right: Minimum of 15 in (38.1 cm)
    • Top: Minimum of 22 in (56 cm)
    • Front: Minimum of 50 in (127 cm)

Waste requirements

  • Solid waste (Tips, RVs, Reagent Packs) collected in waste container. (No manual removal required for operation)
  • Tip trays, substrate bottles and caps stored separately for recycling capability purposes
  • Floor drain only
  • Drain Location:
    • Maximum distance 33 ft (10 m) from the system
    • Maximum height 4 ft (1.5 m) from the floor

Power and environmental requirements

  • Power Supply: 200-240 VAC at 1600 VA, at either 50 Hz or 60 Hz, single phase
  • Frequency: 47–63 Hz
  • Power: No greater than 2,700 W
  • Computer:
    • CPU Intel Core i7-6700 processor
    • Memory: 2 x 16GB DDR4-2133
    • Storage: 2 x 240GB SSD
    • Operating system: Windows 10
    • External communication types
  • RS232 serial interface:
    • Ethernet: LAN using TCP/IP
  • Monitor:
    • Touchscreen display: 21.5 in diagonal TFT LCD with projected capacitive touch
  • Interface:
    • Pixels: 1920 x 1080 RGB
  • Heat output: 4,231 BTU/hr
  • Ambient operating environment
    • Ambient temperature: 64°F to 86°F (18°C to 30°C)
    • Humidity: 20% to 80%
  • Altitude: 7,500 ft (2.286 km)
  • Noise Level: <60 dB at 1000Hz
  • Water: No water requirement
  • Drain Volume: >1.85 gal/hr (>7 L/hr)

DxI 9000 Access Immunoassay Analyzer