Analytical method
|
- Chemiluminescent detector: luminometer
|
Barcoded reagents
|
- Automatic tracking of:
- Number of tests
- Available tests
- Expiration date
- Lot number
- Calibration expiration and printouts
|
Calibration
|
- Calibration curve stability up to 64 days (assay dependent)
- Calibration curves and parameters displayed on screen
|
Communication modes
|
- Unidirectional, bi-directional, bi-directional with true host query
- Both RS-232 and LAN connections are supported
|
Compartment temperatures
|
- Incubator and wash/read wheel: 98.6°F (37°C)
- Sample wheel: 40.1°F to 57.2°F (4.5°C to 14°C)
- Reagent compartment: 39.2°F to 50°F (4°C to 10°C)
|
Immunoassay menu capacity
|
- >50 preprogrammed, barcoded immunoassay methods currently available
|
Measurement principles
|
- Acridan based chemiluminescent Lumi-Phos PRO
|
Sample container sizes
|
- Ability to accommodate any tube between 12–16mm in diameter and 75–100mm in height
- Sample cups:
- 2.0 mL insert cups
- Pediatric insert
- Auto aliquot tube
|
Sample management/capacity
|
- 140 sample tubes (20 racks of 7 tubes), unique continuous loading feature
- Primary tube released after aliquot is made (approx. 1 minute) 300 aliquots (≤200 μL) per hour
- Capable of automation connection
- Barcode symbology:
- Ability to read 1-D and 2-D symbologies with camera technology.
- Sample volume:
- 2-100 μL (assay dependent)
- 500 μL in aliquot vessel, maximum (including any potential reflex tests)
- Sample and reagent pipettor obstruction detection
|
Sample types (assay dependent, partial list)
|
- Serum
- Plasma
- Urine
- Amniotic fluid
- Whole blood
|
Reagent storage capacity
|
- 50 reagent packs:
- System menu can be configured to match laboratory testing patterns (ability to load one pack of 50 different assays or 50 packs of one assay)
- Automatic disposal of empty packs
|
Throughput
|
- Up to 450 tests/hr maximum (one-step assays)
|